SETPOINT OUTPUT TIME-OUT
COUNTER AUTO RESET
SPn tOUt ?
? 001.00
0.01
to
599.99
seconds
SPn AUtO
?
?
NO
NO
@2Ero-Str
2Ero-End
CtLd-Str
CtLd-End
This parameter is only active if the Setpoint Action is set to time out ( t-OUt ).
Enter the value in seconds that the Setpoint output will be active, once the
Setpoint Value is reached.
SETPOINT VALUE
This parameter automatically resets the counter to which the setpoint is
assigned ( SPn ASN ) each time the setpoint value is reached. The automatic reset
can occur at output start, or at output end if the Setpoint Output Action is
programmed for timed output mode. The Reset-to-Count Load selections
(“ CtLd- ”) only apply to Counter A assignment. This reset may be different from
the Counter A Reset Action selected in Module 1.
SPn VAL ?
? 00000100
Count A:
Count B:
Rate:
- 9999999 to 99999999
0 to 9999999
0 to 999999
SELECTION ACTION
NO No Auto Reset.
2Ero-Str Reset to Zero at the start of output activation.
CtLd-End Reset to out only). Load value at the end of output activation
?
SP1 OFF2 ?
?
SP2 OFF1 ?
?
Enter the desired Setpoint value. To enter a negative setpoint value,
increment digit 8 to display a “-” sign (Counter A only).
SETPOINT OUTPUT LOGIC
SPn OUt ?
NOr rEV
NOr
Normal ( NOr ) turns the output “on” when activated and “off” when
deactivated. Reverse ( rEV ) turns the output “off” when activated and “on” when
deactivated.
SETPOINT ANNUNCIATOR
SPn LIt ? NOr rEV
? NOr
Normal ( NOr ) displays the setpoint annunciator when the corresponding
output is “on”. Reverse ( rEV ) displays the setpoint annunciator when the output
is “off”.
SETPOINT OUTPUT POWER-UP STATE
CtLd-Str Reset to Count Load value at the start of output activation.
2Ero-End Reset to Zero at the end of output activation (timed out only).
Count
(timed
SETPOINT 1 OUTPUT OFF AT SETPOINT 2 (SP1 Only)
NO Out2-Str Out2-End
NO
This parameter will deactivate Setpoint 1 output at the Start or End of
Setpoint 2 output (O1 off at O2). The “ -End ” setting only applies if Setpoint 2
Output Action is programmed for timed output.
SETPOINT 2 OUTPUT OFF AT SETPOINT 1 (SP2 Only)
NO Out1-Str Out1-End
NO
This parameter will deactivate Setpoint 2 output at the Start or End of
Setpoint 1 output (O2 off at O1). The “ -End ” setting only applies if Setpoint 1
Output Action is programmed for timed output.
SPn P-UP
?
OFF
ON
SAVE
SETPOINT OUTPUT RESET WITH MANUAL RESET
?
?
SPn StbY ?
?
? OFF
SAVE will restore the output to the same state it was at before the meter was
powered down. ON will activate the output at power up. OFF will deactivate the
output at power up.
SETPOINT BOUNDARY TYPE
SPn tYPE ? HI-ACt LO-ACt
? HI-ACt
High Acting Boundary Type activates the output when the assigned display
value ( SPn ASN ) equals or exceeds the Setpoint value. Low Acting activates the
output when the assigned display value is less than or equal to the Setpoint.
SETPOINT STANDBY OPERATION
NO YES
NO
This parameter only applies to Low Acting Boundary Type setpoints. Select
YES to disable a Low Acting Setpoint at power-up, until the assigned display
value crosses into the output “off” area. Once in the output “off” area, the
Setpoint will then function per the description for Low Acting Boundary Type.
12
SPn rSt ?
YES NO
YES
Selecting YES causes the Setpoint output to deactivate (reset) when the
Setpoint Assigned Counter is reset. The counter reset can occur by the RST
button, User Input, Counter Reset at Power-up or a serial Reset Counter
command.
This output reset will not occur when the Assigned Counter is reset by a
Setpoint generated Counter Auto Reset.
CHANGE DISPLAY COLOR WITH OUTPUT STATE
SPn ChC ?
NO YES
NO
This parameter enables the backlight CUB5 to switch the backlight color when
the output state changes. This parameter is only active for the backlight version.